HeartStart FR3


Product number:


The HeartStart FR3 is Philips best professional-grade AED with advanced features for the professional responder. The FR3 reduces deployment time by eliminating steps and helps you start the right therapy on your patient faster.

Specifications Show all

Physical dimensions
Size with external paddles
12.4" (W) x 7.7" (D) x 13.4" (H) (313 mm x 195 mm x 340 mm) "/mm
13.2 lbs. (6 kg): base unit with 1 battery/pads and pads cable.
13 lbs. (5.9 kg) with optional 75 mm strip chart printer.
Paddle tray and external standard paddles add less than 2.5 lbs. (1.1 kg).
Carrying case adds 4.1 lbs. (1.86 kg). lbs/kg
Truncated Exponential Biphasic.Waveform parameters adjusted as a function of patient impedance.
Output Energy
Manual (selected): 1-10/15/20/30/50/70/100/120/150/170/200 Joules into a 50 Ohm load
AED Mode (single energy output): 150 Joules into a 50 ohm load. Joules/Ohm
Charge Time
Less than 5 seconds to 200 Joules with a new/ fully charged lithium ion battery at 25º C
Shock Delivery
Via multifunction defib electrode pads or paddles
Shock-to-Shock Cycle Time
Typically less than 20 seconds seconds
Minimum Patient Impedance Range
15 Ohm (internal defibrillation); 25 Ohm (external defibrillation) Ohm
Maximum Patient Impedance Range
180 Ohm Ohm
AED Mode
Shock advisory sensitivity and specificity meet AAMI DF-39 guidelines
Physical dimensions Defibrillator model HeartStart MRx (M3536A)
Size without external paddles
12.4" (W) x 7.7" (D) x 11.7" (H) (313 mm x 195 mm x 295 mm) "/mm
CPR feedback and measurement
Compression Depth
Target: -1.50 to -2.00" (-38 to -51 mm) "/mm
Compression Rate
Target: 90 to 120 cpm cpm
Ventilation Volume
Graphic indicator: empty/one-third full/two-thirds full/full
Ventilation Rate(a)
Target for breaths delivered before 60 seconds since last compression: 6 to 16 vpm seconds/vpm
Ventilation Rate(b)
Target for breaths delivered beyond 60 seconds since last compression: 9 to 16 vpm seconds/vpm
SpO2 pulse oximetry
0 to 100% %
Alarm Range(a)
Low Limit: 50 to 99% (Adult/Pediatric) %
Alarm Range(b)
High Limit: 51 to 100% (Adult/Pediatric) %
Alarm Delay
10 seconds
ECG and arrhythmia monitoring
Up to 4 ECG waves displayed and up to 3 ECG waves print simultaneously.
Lead I, II, or III obtained through 3-lead ECG cable and separate monitoring electrodes.
With 5-lead cable, obtain leads I/II/III/aVR/aVL/ aVF or V.
Pads ECG obtained through 2 multifunction defibrillation electrode pads.
Lead Fault
‘Lead Off ’ message and dashed line displayed if an electrode or lead wire becomes disconnected.
Pads Fault
Dashed line displayed if a pad becomes disconnected.
Heart Rate Display
Digital readout on display 15 to 300 bpm with accuracy ±10% bpm/%
Heart Rate/Arrhythmia Alarms
HR/Asystole/VFIB/VTACH/VTACH/extreme tachycardia/extreme bradycardia/PVC rate.
ECG Size
2.5/5/10/20/40 mm/mV autogain mm/mV
Noninvasive Blood Pressure
Pressure Range(a)
Systolic: 40 to 260 mmHg mmHg
Pressure Range(b)
Diastolic: 20 to 200 mmHg mmHg
Initial Pressure
Adult: 160 mmHg and Pediatric: 120 mmHg mmHg
Maximum Pressure
280 mmHg mmHg
Alarm Range(d)
Systolic high limit: 30 - 270 (Adult) and 35 - 180 (Pediatric)
Alarm Range(e)
Systolic low limit: 30 - 265 (Adult) and 30 - 175 (Pediatric)
Alarm Range(f)
Diastolic high limit: 18 - 240 (Adult) and 18 - 150 (Pediatric)
Alarm Range(g)
Diastolic low limit: 10 - 240 (Adult) and 10 - 145 (Pediatric)
2.7" high x 5.3" wide x 8.7" deep (6.9 cm x 13.5 cm x 22.1 cm)
3 lbs 8 oz (1.6 kg) with FR3 standard battery installed
Environmental & physical requirements
Solids/Water resistance
Operating temperature
0º - 45º C (32º - 113º F) ° C/° F
Storage temperature
-20 - 70º C (-4º - 158º F) ° C/° F
Operating humidity
0% to 95% relative %
Operating altitude
0 to 15000 ft (0 to 4500 m) ft/m
Storage altitude
0 to 15000 ft (0 to 4500 m) ft/m
Mechanical shock
Bump: IEC 68-2-29
Freefall: IEC 68-2-32
Operating vibration
MIL STD 810E 514.4 Category 6 Helicopter
General storage vibration
UH60 Non-Operating: IEC 68-2-6 Swept Sine Vibration and IEC 68-2-64 Random Vibration
Meets EN 60601-1, UL 2601-1/CSA C22.2 No. 601-1
Strip Chart Recorder
Standard: 50 mm (paper width) thermal array printer mm
Optional: 75 mm (paper width) thermal array printer mm
Continuous ECG Strip
Prints primary ECG lead with event annotations and measurements in real-time or with 10-second delay
Auto Printing
Recorder can be configured to print marked events/charge/ shock and alarms
Event Summary, 12-Lead/Operational Check/ Configuration/Status Log and Device Information
Paper Size
1.97” (50 mm) W by 100 ft. (30 m) L and 2.95” (75 mm) W by 100 ft. (30 m) L "/ft/mm/m
6.3 Ah, 14.8 V, rechargeable lithium ion Ah/V
6.5" (H) x 3.8" (W) x 1.6" (D) (165 mm x 95 mm x 42 mm) "/mm
1.6 lb. (0.73 kg) lb/kg
Charge Time
Approximately 3 hours from fully depleted to 100%. 90 minutes from fully depleted to 80% hours/%
At least 5 hours of continuous 12-lead ECG/SpO2 and CO2 monitoring.
With NBP every 15 minutes on one new fully charged battery. hours
At least 3.5 hours of continuous 12-lead ECG/SpO2 and CO2 monitoring.
With NBP every 15 minutes and pacing at 180 ppm at 160 mA on one new fully charged battery. hours
Meets IEC529 class IP55 with battery installed
Operating/standby: 32°–122°F (0°– 50°C)
0 –15,000 feet (0 –4,572 meters)
8.4" diagonal (128 mm x 171 mm) "/mm
TFT color LCD
480 x 640 pixels (VGA) VGA
Wave Viewing Time
5 seconds (ECG) seconds
12-Lead ECG
12-Lead cable: leads I/II/III/aVR/aVL/ aVF and V/C1-V/C6
Display View
All 12-lead ECG waves display simultaneously
Strip Recorder
All 12-leads print on the strip chart printer in 3x4 format
CompactFlash data card; Bluetooth® dial-up Internet connection, Bluetooth® FTP
End-tidal CO2
0 to 99 mmHg mmHg
1 mmHg (0.1 kPa) mmHg/kPa
Sample Size
50 ml per minute ml
Alarm Range(h)
Low Limit: 10 to 95 mmHg (Adult/Pediatric) mmHg
Alarm Range(i)
High Limit: 20 to 100 mmHg (Adult/Pediatric) mmHg
Data storage
12 hours of continuous ECG waveforms and events. Plus 50 12-lead ECG reports
Data Card
12 hours of continuous ECG waveforms and events.
Plus 50 12-lead ECG reports on a CompactFlash memory card
ECG display
LCD color display, 320 x 240 pixels. 2.8" x 2.1" (7.2 cm x 5.4 cm)
1 Hz to 30 Hz (-3dB), nominal (non-diagnostic)
Monitored Lead
Lead II using anterior-anterior adult pads placement
Meets MIL-STD-810F 516.4, Procedure IV
Patient Impedance Range
"Minimum: 15 ohm (internal defibrillation); 25 ohm (external defibrillation) Maximum: 180 ohm"
Meets MIL-STD-810F 514.5 C-17
Abuse tolerance
After a ibe-meter drop to any edge, corner, or surface in standby mode
AED Mode
Shock advisory sensitivity and specificity meet AAMI DF-39 guidelines